5.2.5. Resistance / Continuity test

Selector switch
Ranges
position
500 Ω
500 Ω
5 kΩ
50 kΩ
Ω
500 kΩ
5 MΩ
50 MΩ
(∗) Overload protection can be reset automatically.
Number of count : 5000
Range selection : automatic or manual (fixed in continuity mode)
Maximum open circuit voltage : 4 V
Detection threshold in continuity mode : 10 Ω to 20 Ω
Response time in continuity mode : < 10 ms
For measurements in 5 MΩ and 50 MΩ ranges, the use of shielded connection is highly
recommended.

5.2.6. Capacitance

Note
Discharge all capacitors before taking measurements.
Selector switch
Ranges
Accuracy
position
50 nF
500 nF
5 µF
1 % R + 4 D
50 µF
500 µF
5000 µF
50 mF
(∗) Overload protection can be reset automatically.
To reduce the influence of the ambient electromagnetic fields on the measurement, it is
recommended to use shielded cables for measurements in 50 nF range.
Number of count : 5000
Range selection : automatic or manual
Maximum open circuit voltage : 4 V

5.2.7. Diode threshold voltage measurement

Measurable voltages : 0 to 1.999 V
Measurement current : 1 mA ± 20 %
Resolution : 1 mV
Protection : 600 V
, can be reset automatically
RMS
OL : displays at 4 V
The battery can be tested without leads if : 03.8 V < V battery < OL
28
Measurement
Accuracy
current
0.3 % R + 3 D
1 mA
1 mA
100 µA
0.3 % R + 3 D
10 µA
1 µA
0.5 % R + 3 D
100 nA
1 % R + 5 D
50 nA
Measurement
Max measurement
current
100 nA
1 µA
10 µA
100 µA
1 mA
≈ 3 s/mF
1 mA
≈ 3 s/mF
1 mA
Chapter II
Protection
Resolution
(∗)
0.1 Ω
0.1 Ω
1 Ω
10 Ω
600 V
RMS
100 Ω
1 kΩ
10 kΩ
Protection
Resolution
(∗)
time
< 1 s
10 pF
< 1 s
100 pF
< 1 s
1 nF
600 V
< 1 s
10 nF
RMS
< 2 s
100 nF
1 µF
10 µF
Portable digital multimeters