Resistance / Continuity; Capacitance; Diode Threshold Voltage Measurement - Metrix MX 553 Manual De Instrucciones

Polimetro de sobremesa
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

5.5.

Resistance / Continuity

Ranges
500 Ω/€
5 kΩ
50 kΩ
500 kΩ
5 MΩ∗∗
50 MΩ∗∗
* Overload protection can be reset automatically.
**It is highly recommended to use very short and shielded test leads for measurements in
this range (>1MΩ).
When performing resistance measurements, connect the COM terminal to the earth.
Number of points:
Range selection:
Maximum open circuit voltage:
Detection threshold in continuity mode:
Response time in continuity mode:
5.6.

Capacitance

"
Note
Discharge all capacitors before taking measurements.
Ranges
50 nF**
1.5% R + 2 D
500 nF
5 µF
50 µF
500 µF
5 000µF
50 mF
* Overload protection can be reset automatically.
**It is highly recommended to use very short and shielded test leads for measurements in
this range.
Number of points:
Range selection:
Maximum open circuit voltage:
5.7.

Diode threshold voltage measurement

Measurable voltages:
Measurement current:
Resolution:
Protection:
26
Accuracy
0.1% R + 5 D
0.1% R + 3 D
0.1% R + 3 D
0.1% R + 3 D
0.3% R + 3 D
1% R + 3 D
Accuracy
Measurement
current
100 nA
1% R + 2 D
1 µA
1% R + 2 D
10 µA
1% R + 2 D
100 µA
1% R + 2 D
1 mA
2% R + 2 D
1 mA
2% R + 2 D
1 mA
Measurement
current
1 mA
100 µA
10 µA
1 µA
100 nA
10 nA
50 000 (or 5 000 - refer to § 3.3.)
Automatic or manual
(fixed in continuity mode)
7 V
10 Ω to 20 Ω
1 ms
Max measurement
time
0.5 s
0.5 s
0.5 s
0.5 s
1.5 s
3 s/mF
3 s/mF
5 000
Automatic or manual
7 V
0 to 2 V
1 mA ± 20%
1 mV
600 V
, can be reset automatically
RMS
Protection*
Resolution
600 V
10 mΩ
RMS
600 V
100 mΩ
RMS
600 V
RMS
600 V
RMS
100 Ω
600 V
RMS
600 V
RMS
Protection*
600 V
RMS
600 V
RMS
600 V
RMS
600 V
RMS
600 V
RMS
600 V
RMS
600 V
RMS
Benchtop multimeter
Chapter II
1 Ω
10 Ω
1 kΩ
Resolution
10 pF
100 pF
1 nF
10 nF
100 nF
1 µF
10 µF

Publicidad

Tabla de contenido
loading

Tabla de contenido