1.5 dTemp, dTemp, Ramp Rate and Pause
eferen
To add dTemp, dTemp, Ramp Rate and Pause to a constant temperature step:
1. In the Edit Program screen, selects a step. Touch Options to open the Step
2. To edit an item, click on the appropriate text box.
3. Enter a value in the pop-up keyboard and then touch Ok.
• dTemp: adds temperature increments to step with each cycle.
• dTime: adds time increments to the holding time within the step for each cycle.
• Ramp Rate: Adjust the heating or cooling ramp rate.
Options window.
Touch +/- in the keyboard to shift between "+" and "–" (to increase or
decrease temperature for each cycle), and then enter the number of degrees of
increment temperature. The allowable increment temperature range is 0.1°C ~
10.0°C.
Note:
Parameters cannot be set that would cause a step to fall outside of the
operating range of the instrument (i.e. below 0°C or above 105°C).
Touch +/- in the keyboard to shift between "+" and "–" (to increase or
decrease time for each cycle), and then enter the seconds. The allowable
increment time range is 1 sec ~ 120 sec.
Note:
Parameters cannot be set that would cause the dwell time to fall below zero.
The default ramping rate is the fastest rate.
To set the Ramp Rate function:
Enter the value of ramp rate in the keyboard. Input range is 0.1 ~ 4.0°C /sec.
Note:
The actual maximum ramp rate of heating and cooling varies with block
module type. Refer to the instrument specifications for the maximum value.
If a value larger than the block's maximum rate is entered, the block's
maximum rate will be used. If you do not want to slow the running, leave this
option empty.
53