Sample Volume And Lid Temperature; Delete Step - LongGene A200 Manual Del Usuario

Termociclador
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

1.8 Sample Volume and Lid Temperature

1.
2. Enter new value in the pop-up keyboard and then touch Ok.
Temperature control mode:
Sim-tube mode: This mode will control the sample temperature to the target
temperature as fast as possible with sample not overheat.
Block mode: This mode will control the block temperature to the target
temperature as fast as possible with block not overheat.
Sim-tube mode is the preferred control method, especially when steps hold time
is 30 seconds or less. It is important to input correct sample volumes for the most
accurate dwell time.
For lid temperature control, it is recommend to select a temperature five to ten
Celsius degrees above the highest temperature of the program steps (e. g., the
pre-denaturation step) to protect sample from evaporation and condensation.

1.9 Delete Step

1. In the Edit Program screen, touch a step to select it.
2. Touch Delete to delete this step.
In the Edit Program screen, touch Volume or Lid text box.
Note:
When a sample volume between 1 and 100µl is entered, the Sim-tube mode
will be used.
When a sample volume of 0µl is entered, the Block mode will be used.
Note:
Setting a lid temperature below ambient will cause the program never start.