Perform Run
1. Load Samples
1. Place PCR tubes or microplate in the sample wells of the sample block.
2. Close the lid and turn the lid knob clockwise (see the 'Tighten' arrow symbol
2. Select Program to Run
To select a program to run,
• From the Home screen, touch New Program, create a new program and then
• Select a RECENT run program from the All Program screen and touch Run.
• Select a saved program from the All Program screen and then touch Run.
• Select a saved program from the All Program screen, touch Edit, change step
1. After you touch Run, the Run Setting window opens.
IMPORTANT!
Do not touch the surface of the sample block or lid surface to avoid
hurt because they may be very hot (105°C for lid)!
Note:
• The cap of tubes or film on microplate must be sealed tightly to prevent
sample evaporation. Tubes or microplate should have heat resistance of up
to about 120°C.
• For best results, use thin walled tubes designed for PCR.
• The sample wells of the sample block should keep clean.
• The height of tubes should be the same so that the lid can keep stable and
gives even pressure on each tube.
on the top surface of the lid knob) until you hear 'click ...' sound.
touch Run.
parameters, and then touch Run.
58