If in doubt, consult the qualified service
personnel.
3. The leakage current could increase
when connected to other equipment.
4. For this particular equipment, all
accessory equipment connected as
noted above, must be connected to
mains via an additional isolation
transformer conforming with the
construction requirements of IEC60601-
1 and providing at least Basic Insulation.
5. This equipment generates, uses, and
can radiate radio frequency energy. If it is
not installed and used in accordance with
the instruction manual, it may cause
interference to other equipment. If this
unit causes interference (which can be
determined by unplugging the power
cord from the unit), try these measures:
Relocate the unit with respect to the
susceptible equipment. Plug this unit and
the susceptible equipment into different
branch circuit.
Consult your dealer. (According to standard
EN60601-1-2 and CISPR11, Class B,
Group 1)