10.1 Error codes

The Incubator has built in fault diagnostics. If a fault occurs, this system will display an error code in
the LED display to help the service engineer rectify the problem. Please see the table below for details
of error codes and other faults. In the unlikely event that a problem does occur, note down which
code / fault you observe and contact Antylia Scientific for further assistance. If you have your own
service personnel available for repair, a comprehensive service manual is available on request.
Fault
Temperature probe PT100
reading low
Temperature probe PT100
reading high
Triac fault
Rotisserie / shaker motor
stalled (INC-200D-H only)
Software crash
11 Specification
(model INC-200D-M & INC-200D-H)
Overall dimensions
Height:
Depth:
Width:
Incubator dimensions
Height:
Depth:
Width:
Weight:
Capacity (nominal):
Temperature range:
Temperature precision:
Temperature fluctuation: ±0.1 °C (@37°C)
Power rating:
300W)
Temperature variation:
Safety system in operation
Automatic over-temp system
activated
Automatic over-temp system
activated
Secondary relay control activated
Cuts power to motor
Cuts mains power
435 mm
380 mm
380 mm
200 mm
230 mm
250 mm
24 Kg
18 litres
Ambient +5°C – 80 °C
±0.5 °C
350W (INC-200D-M -
<0.25°C
Fault displayed by
LED displays Err and 0.1
alternatively
LED displays Err and 0.2
alternatively
Top left dot flashes in display
No movement seen but
Rotisserie red indicator light
flashes
No temperature or motor
power, all LED's blank