R
5 USING THE CONDUCTIVITY METER (continued)
When a range other than AUTO is selected, the meter can be
calibrated only for that particular range. If a sample is measured which
has a higher conductivity level than the selected measuring range, the
error message 'OR' is displayed.
To select the measuring range (Figure 3):
1
With the meter switched off, press and hold the '°C/°F' button,
then switch the meter on. Release the '°C/°F' button.
2
The meter goes into range selection mode. The currently
selected range is shown in the lower display. The upper display
shows the maximum possible reading for the selected range.
Press 'HOLD ENT' repeatedly until the required range is
displayed (PU, LO or HI).
3
The meter automatically conrms the last selection if no button is
pressed for 5 seconds. The upper display momentarily shows
'CO'. The LCD shows the power-up sequence and the meter
goes to measurement mode.
°C/°F
Auto conrmation occurs for the last selected range if no key is pressed for 5 seconds.
en-9
All manuals and user guides at all-guides.com
Figure 3: Selecting the Measuring Range
HOLD
ENT
HOLD
HOLD
ENT
ENT
Power-up
Sequence
www.elcometer.com