Hach DR1300 FL Manual Del Usuario página 25

Ocultar thumbs Ver también para DR1300 FL:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 27
Problem
The instrument
reading is
different from
the online
analyzer
reading.
The instrument
reading is
different from
the online
analyzer
reading.
Section 12 Consumables and replacement parts
Personal injury hazard. Use of non-approved parts may cause
personal injury, damage to the instrument or equipment
malfunction. The replacement parts in this section are approved by
the manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact
the appropriate distributor or refer to the company website for contact
information.
Consumables
Description
DR1300 FL Fluorescence Standards Kit
Reagent kit, ULR Fluorescence Total Chlorine,
includes:
Sample vial holder, two16-mm vials with caps and two
reagent bottles in dropper bottles
Reagent kit, ULR Fluorescence Free Chlorine, includes:
Sample vial holder, two16-mm vials with caps and two
reagent bottles in dropper bottles
Possible cause
The sample is not a
representative sample
and is not well mixed.
The grab sample was
not analyzed
immediately.
Note: The DPD method is
less accurate in the lower
ppb range compared to the
fluorescence method.
Other oxidants do not
interfere with the
fluorescence method.
W A R N I N G
Solution
Make sure to get a representative
sample. Refer to the instructions in
the Sample Collection section of the
test procedure.
Analyze the grab samples
immediately. The chlorine
concentration decreases quickly in
ultra low range chlorine samples.
Item no.
34630000
34252000
34251000
English 25

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Lpv449.97.01002Lpv449.98.01002Lpv449.99.00002

Tabla de contenido