Greenlee DM-70 Manual De Instrucciones página 11

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Accuracy
See the Specifications section for operating conditions and temperature coefficient.
Accuracy is specified as follows: ± (a percentage of the reading + a fixed amount)
at 23 °C ± 5 °C (73.4 °F ± 9 °F), 0% to 80% relative humidity.
DC Current
Range
320.0 µA
± (1.0% + 0.2 µA)
3200 µA
AC Voltage
Range
± (1.5% + 0.005 V)
3.200 V
± (1.5% + 0.05 V)
32.00 V
320.0 V
600 V
DC Voltage
Range
± (0.5% + 0.2 mV)
320.0 mV
± (0.5% + 0.002 V)
3.200 V
± (0.5% + 0.02 V)
32.00 V
320.0 V
600 V
Resistance
Range
320.0 Ω
± (1.0% + 0.4 Ω)
± (0.8% + 0.002 kΩ)
3.200 kΩ
± (0.8% + 0.02 kΩ)
32.00 kΩ
± (0.8% + 0.2 kΩ)
320.0 kΩ
± (0.8% + 0.002 MΩ)
3.200 MΩ
± (2.0% + 0.05 MΩ)
32.00 MΩ
Diode Test
Measuring Range: 3.200 V
Test Current: < 1.5 mA
Open Circuit Voltage: < 3.3 VDC
Continuity
Tone indicates continuity. The threshold is approximately 15 Ω.
All manuals and user guides at all-guides.com
Accuracy
± (1.0% + 2 µA)
Accuracy
± (1.5% + 0.5 V)
± (1.5% + 5 V)
Accuracy
± (0.5% + 0.2 V)
± (0.5% + 2 V)
Accuracy
Burden Voltage
3 mV/µA max.
Frequency Range
40 Hz to 300 Hz
40 Hz to 500 Hz
Input Impedance
10 MΩ
Open Circuit Voltage
Approximately
1.3 VDC
DM-70
Input Impedance
10 MΩ,
100 pF nominal
11

Publicidad

Tabla de contenido
loading

Tabla de contenido