Chapter 4 - Technical Specification
IEC60601 compliance test
Immunity test
level
Electrical fast
±2 kV for power supply lines
transient/burst
±1 kV for input / output lines
IEC61000-4-4
Surge
±1 kV line to line
IEC61000-4-5
±2 kV line to earth
<5% U
(>95%dip in U
for 0.5 cycle
Voltage dips, short
40% U
interruptions and
(60% dip in U
voltage variations
for 5 cycles
on power supply
lines
IEC61000-4-11
70% U
(30% dip in U
for 25 cycles
<5% U
(>95% dip in U
for 5 seconds
Power frequency
(50/60Hz)
3 A/m
Magnetic field
IEC61000-4-8
Note: U T is the a.c. mains voltage prior to the application of the test level.
Conductive RF
3Vrms
IEC61000-4-6
150kHz to 80MHz
Page 68 of 76
English
Electromagnetic environment-guidance
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
T
)
T
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
T
)
device requires continued operation during mains
T
interruptions, it is recommended that the device be
powered from an uninterruptible power supply.
T
)
T
If the user of the device requires continued operation
T
)
during mains interruptions, it is recommended that the
T
device be powered from an uninterruptible power supply.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
Portable and mobile RF communications equipment
should be used no closer to any part of the device,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
d=1.2√P
Part Number: G108707/8GCA
Copyright 2018, Optos plc. All rights reserved.