Aesculap
High frequency surgical device GN060
3.3
Moving stacks of units
►
Always lift stacks from underneath the bottommost unit.
3.4
Dismantling stacks of devices
Dismounting a stack of products fitted with stacking devices:
Fig. 5
Dismantling stacks of devices
Legend
B Locking screw
►
Apply a screwdriver on locking screw B and turn it counterclockwise to
the positive stop.
►
Remove the topmost device.
►
Screw back on the feet of the device, either by hand or using a screw-
driver.
3.5
Presetting the correct voltage
The mains voltage must correspond to the voltage indicated on the type
plate at the back of the unit.
The correct voltage for the control unit is indicated in the window located
at the back of the device. The standard voltage for the unit in Europe is:
230 V.
Risk of fatal injury from electric shock!
►
Unplug the device before replacing the fuse sets.
DANGER
►
If the indicated voltage does not correspond to the actual mains volt-
age, change the setting at the rear panel of the power unit:
– Release clip B using a suitable screwdriver A and remove the fuse
holder 25.
– Pull out the voltage selection element 24 under the fuse holder.
– Reinsert the voltage switching element so that the desired
voltage C is legible.
– Reinsert the fuse holder 25.
8
®
B
Voltage range
100 V to 120 V (+/- 10 %)
220 V to 240 V (+/- 10 %)
23
A
B
25
Fig. 6
Device connection with fuse holder and voltage selection ele-
ment
Legend
23 Device connector
24 Voltage selection element
25 Fuse holder
A Screwdriver
B Clip
C Voltage switching element imprint
Imprint C
110
230
24
C