Full HD CMOS camera system PV470
Unplug the device by pulling on the plug, and never on the power cord.
In order to avoid damage to the electrical components and inoperabil-
ity:
– Do not bend the product.
– Avoid bending, torsional, tensile or compression loading.
Operate the product with original manufacturer's parts only: System
cable, video cable and power cord.
The Full HD CMOS camera system is intended for daily to weekly use.
Note
The connection of electrical devices to a multiple socket creates a system
and can lead to a reduced level of safety.
Note
The camera case is only transport packaging.
3.

Product description

3.1

Scope of supply

Camera control unit
Art. no.
Designation
PV470
Full HD CMOS camera control unit
DVI-cable
TA013748
Instructions for use
Camera head
Art. no.
Designation
PV471
Pendulum camera head
–or–
PV472
Zoom camera head
–or–
PV473
Fixed focal length camera head
3.2

Components required for operation

In addition to the camera system, the following components are required
for an endoscopic intervention:
Power cord
Endoscope
Light source
Light cable
Monitor
4
®
3.3

Intended use

The Full HD CMOS camera system PV470/PV471/PV472/PV473 is used in
medical endoscopy, together with an endoscope, a light source and a
monitor.
It serves to visualize the endoscopic image for display on a color monitor.
Indication
The product is indicated for supporting diagnosis and treatment in endo-
scopic imaging.
Contraindication
Due to the universal suitability of the product for endoscopy, no specific
contraindications for its use can be given. The use of the product is con-
traindicated if the endoscopic procedure are contraindicated or if its use
could pose a risk to the patient.
3.4

Operating principle

The Endoscope visualizes the operation site and the endoscopic lens
focuses this image onto the CMOS sensor, where it is converted into elec-
trical signals that can be further amplified. The image brightness is auto-
matically adjusted to the light intensity (autoshutter).
Through camera cable 26, the image signals are transferred to camera
control unit, where they are further amplified for display on the monitor.
The video signal processed in this way can then be picked up at the video
output sockets for connecting monitors, video printers or any other digital
archiving systems.
For areas that are particularly difficult to illuminate, the camera system
offers a dynamic electronic brightness enhancement function (GAIN),
which automatically adapts to different light conditions, and a choice of
mode settings for different situations during application.
The video camera system is also equipped with a manual white balance
system for optimal color rendition.
Connections on the front panel of the camera control unit
Control elements / indicators
"Power off" switch 1
"Power on" switch 2
Pushbutton white balance 4
Camera socket 5
Description
To switch the device on and off.
LED display 3 above the "Power
on" switch 2 lights up when the
device is switched on.
For precise white-balancing,
ensuring realistic rendition of all
other colors
For connecting the camera plug at
the end of the camera cable