Other Malfunctions/Possible Faults - MIETHKE 75100000 Instrucciones De Uso

Ocultar thumbs Ver también para 75100000:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Reader Unit Set for SENSOR RESERVOIR
Message displayed
System error
- Calibration data incorrect
System error
- contact Service
Keypad faulty
Temperature increase out of
range
Temperature value out of range
Timeout!
Restart measurement!
Ambient pressure incorrect
:
Ambient pressure from 800 to
1.100 mbar valid

OTHER MALFUNCTIONS/POSSIBLE FAULTS

Message displayed
Device cannot be activated
Device switches off
Cause
Calibration data in implant cor-
rupted or not readable (only
applies if no SD card has been
inserted)
System test detected an error
A pressed key is registered upon
activation of the reader unit
A
spontaneous
temperature
increase
in
implant
2 K or beyond 39.0 °C
Temperature data measured from
implant not plausible
- physiologically meaningless
data
The
time
between
measurement and successfully
establishing communication was
exceeded (60 seconds).
The permissible barometric pres-
sure during measurement is too
low or too high.
Cause
Battery is completely empty
Unfavourable operating condi-
tions (e.g. low humidity or unsu-
itable floor coverings)
INSTRUCTIONS FOR USE |
Fault detection/rectification
Measurement is stopped. Deactivate and re-
activate device. If the error persists, contact
Technical Service.
Measurement with inserted SD card possible.
Device function locked. Deactivate and re-
activate device. If the error persists, contact
Technical Service.
Release the key.
Deactivate and reactivate device.
Measurement is stopped. Leave to rest for 10
above
minutes. If the error persists, contact Tech-
nical Service.
Measurement is stopped. Deactivate and re-
activate device. If the error persists, contact
Technical Service.
starting
Restart measurement.
Optimize distance between antenna and im-
plant.
The reader unit can only be used at barome-
tric pressures from 800 to 1100 hPa.
Current measurement is stopped.
Fault detection/rectification
Connect to power supply.
It takes approx. 6 hours to charge the battery
fully. The reader unit can also be operated
during the charging process (with connected
power supply unit).
Note: Charging is not possible at an ambient
temperature of more than 35°C.
Switch device back on.
If the error persists, contact Technical
Service.