Malfunctions Displayed With An Error Message - MIETHKE 75100000 Instrucciones De Uso

Ocultar thumbs Ver también para 75100000:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

| INSTRUCTIONS FOR USE
GB

MALFUNCTIONS DISPLAYED WITH AN ERROR MESSAGE

Message displayed
Battery flat - auto off
Battery voltage incompatible -
use original power supply
Low Battery Voltage
Flashing of crossed-out
antenna symbol
:
Antenna error
Flashing of crossed-out
antenna symbol
:
Antenna not connected
Flashing of crossed-out
antenna symbol
:
No communication
Flashing of crossed-out anten-
na symbol
:
SD card has been removed.
Measurement possible
Data error!
:
File opening not possible
Continuous key activation
Keypad error
Pressure values out of range
Input voltage incompatible.
50
Cause
Battery capacity
exhausted (0 %)
Battery voltage of the reader unit
is too low
Battery voltage is too low
Antenna faulty
Antenna not connected at the
start of measurement
- or -
Antenna was disconnected duri-
ng measurement
Data recording interrupted du-
ring continuous measurement
(telemetric link disrupted)
SD card was removed during
continuous measurement
Validation of file failed
Persistent activation of a key
> 60 seconds
Pressure data measured from
implant not plausible
- physiologically meaningless
data
Power supply voltage is too high
Reader Unit Set for SENSOR RESERVOIR
Fault detection/rectification
All data are saved after 2 min. Reader unit
switches off automatically. Connect to original
power supply.
Reader unit switches off automatically after
20 seconds. Connect to original power sup-
ply.
Backlight switches off after 3 seconds. Con-
nect to power supply. No interruption to on-
going measurements.
Deactivate and reactivate device. If the error
persists, contact Technical Service
Connect antenna.
Measurement starts again.
-or-
Connect antenna. Measurement continues.
Measurement resumes automatically once
communication is restored.
Insert SD card. Start measurement again.
File cannot be opened. Try again.
Release the key.
Measurement continues. If the error persists,
contact Technical Service.
Reader unit switches off automatically after
20 seconds.
Use original power supply.

Publicidad

Tabla de contenido
loading

Tabla de contenido