INDICATIONS FOR USE (continued)
STERRAD
NX
Standard Cycle, Wrap & Aesculap
®
®
InstruSafe® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a
healthcare provider. InstruSafe Instrument Protection System cassettes are intended to allow sterilization of the enclosed medical
devices during a Sterrad NX Standard Sterilization Cycle. The InstruSafe Instrument Protection System cassettes are intended to be
used in conjunction with a legally marketed wrap or Aesculap rigid container. The InstruSafe Instrument Protection System cassettes are
not intended on their own to maintain sterility. A full list of device models is provided in Appendix A.
Sterilization Methods and configurations
- Sterrad NX Standard Sterilization Cycle
Lumen Claims for STERRAD NX Standard Sterilization Cycle
Minimum Inside
Summit Cassette Model
Diameter
IN-2681
1mm
IN-2681
3mm
IN-8987-S
1mm
IN-8615
2.3mm
IN-6105
4mm
Note: The worst case validated load based on vent-to-volume calculation
is the IN-2681 tray.
STERRAD
100NX
Express Cycle, Wrap
®
®
InstruSafe® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a
healthcare provider. InstruSafe Instrument Protection System cassettes are intended to allow sterilization of the enclosed medical
devices during a Sterrad® 100NX Express Cycle. The InstruSafe Instrument Protection System cassettes are intended to be used in
conjunction with a legally marketed wrap. The InstruSafe Instrument Protection System cassettes are not intended on their own to
maintain sterility. The InstruSafe Instrument Protection System has no lumen claims for the Sterrad 100NX Express Cycle.
STERRAD
100NX
Flex Sterilization Cycle, Wrap & Aesculap
®
®
InstruSafe® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a
healthcare provider. InstruSafe Instrument Protection System cassettes are intended to allow sterilization of the enclosed medical
devices during a Sterrad 100NX Flex Sterilization Cycle. The InstruSafe Instrument Protection System cassettes are intended to be used
in conjunction with a legally marketed wrap or Aesculap rigid container. The InstruSafe Instrument Protection System cassettes are not
intended on their own to maintain sterility. A full list of device models is provided in Appendix A.
STERRAD 100NX Flex Sterilization Cycle
Summit Cassette Model
IN-0000
IN-6105
* Validated by Summit Medical for use in STERRAD 100NX Flex
Sterilization Cycle ONLY. Consult container instructions to ensure that
contents do not exceed the sterilization container's intended load
claims.
Lumen Claims for STERRAD 100NX Flex Sterilization Cycle
Summit
Minimum
Cassette
Inside
Model
Diameter
IN-0000
1mm
IN-8823
1mm
IN-7344
1mm
IN-6105
4mm
®
Maximum
Number of
Length
Lumens
65mm
1
200mm
1
500mm
5
210mm
5
235mm
1
Aesculap Container Model
*JM440
*JM440
Maximum
Number of
Wrap/Rigid
Length
Lumens
Container
Wrap + Rigid
850mm
1
Container
Wrap + Rigid
850mm
1
Container
850mm
1
Wrap
Wrap + Rigid
235mm
1
Container
Rigid Container
Rigid Container
®
The worst case validated load based on vent-to-volume calculation is
the IN-0000 tray.
Note: The IN-0000 tray is for testing purposes only.