ASSEMBLY

1. Place the instruments in the predetermined holders or area defined by the locating posts so that all instrument surfaces will be
exposed to sterilant. See Figure A. Be sure that only one instrument is in each slot. When possible, disassemble or open all parts
of the instrumentation per the instrument manufacturer's instructions.
DO NOT overload holders or exceed weight limits of cassettes. See Appendix A.
Figure A.
2. Ensure that handles are positioned inside the cassette (retracted position). Shown in Figure B.
Figure B.
3. Place the cassette cover on the cassette base ensuring that the latches are properly positioned with the latch slot on the cassette
cover, then close the latches. See Figure C.
Figure C.
Figure D.
Improper Alignment
Unlatch
Proper Alignment
Note: Instruments (e.g. endoscopes and instruments with lumens or
channels) should be prepared and sterilized according to the
instrument manufacturer's instructions for use.
Note: The latches are offset so that the cassette cover always aligns
properly with the base. This ensures that the holders are aligned
when the cover is secured. See Figure D for proper alignment.
DO NOT force the cassette cover onto the cassette base as it may
damage instruments and the cassette.
5