Cables and accessories
Motor Cord
Foot Control Cord
Power Cord
Recommended separation distances between portable and mobile RF communications equipment and the Primado2.
The Primado 2 is intended for use in an electromagnetic environment in which radiated RF disturbances are control LCD. The customer or the user of the
Primado 2 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the Primado 2 as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance (d) in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where (P) is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and
people.
34
Maximum length
Complies with
RF emissions, CISPR11:
Harmonic emissions (exception 120V):
3.5m(Unshielded)
Voltage fluctuations/ flicker emission (exception 120V):
Electrostatic discharge ( ESD ):
3.5m(Unshielded)
Electric fast transient / burst:
Surge:
3.6m(Unshielded)
Voltage dips, short interruptions and voltage variations on
power supply input lines:
Power frequency (50/60Hz) magnetic field:
Conducted RF:
Radiated RF:
Separation distance according to frequency of transmitter
150kHz to 80MHz
d=1.2 P
0.12
0.38
1.2
3.8
12
m
80MHz to 800MHz
d=1.2 P
0.12
0.38
1.2
3.8
12
Class B/Group 1
EN/IEC61000-3-2
EN/IEC61000-3-3
EN/IEC61000-4-2
EN/IEC61000-4-4
EN/IEC61000-4-5
EN/IEC61000-4-11
EN/IEC61000-4-8
EN/IEC61000-4-6
EN/IEC61000-4-3
800MHz to 2.5GHz
d= 2.3 P
0.23
0.73
2.3
7.3
23