EN – ENGLISH
Drying
• Products should be dry at this point. If wetness or excess liquid is detected, dry with a soft clean towel.
• Medically compressed air can be used if needed.
Optional Automatic Wash / Disinfect
CAUTIONS
• Alkaline and highly acidic detergents and solutions cause damage to the devices.
• Channels and crevices found on these devices require particular attention during cleaning.
• Pay special attention to the water quality used throughout reprocessing. Hard water can damage the surface of the equipment. Avoid
using hard water. Instead use purified water unless otherwise specified.
Limitations on Reprocessing
• Repeated processing has negative effects on these devices and is not recommended for routine use.
• It is important to have Integra NeuroSpecialists perform routine inspections (twice yearly is recommended).
See contact information.
INSTRUCTIONS
Cleaning - Automated Equipment: Neutral pH Detergent
Method
1.
Prepare equipment for cleaning by disassembling removable parts and loosening connections.
2. Rinse equipment in warm water before placing into washer.
3. Load device into the washer and place small parts in container or tray inside the washer unit in order to avoid losing small
components. NOTE: Load devices carefully into washer in order to avoid collision.
4. Follow the instructions listed below and set washer machine to these exact parameters:
Phase
Pre-wash 1
Enzyme Wash
Wash 1
Rinse 1
Thermal Rinse**
** Optional phase for disinfection of components – minimum water temperature as indicated or per worker manufacturer
specifications for the thermal rinse cycle.
NOTE: Any deviation from this guideline could result in damage to the equipment as well as improper cleaning results.
Rinse with purified water. Do not perform if parameters cannot be achieved.
5. Remove from washer and dry completely if needed.
6. Inspect equipment to make sure there is no visible organic debris or residue from cleaning agent.
Repeat process if any visible soil is detected.
The life expectancy of the MAYFIELD products is expected to be as long as 7 years.
Care and Storage
The MAYFIELD Pediatric Radiolucent Horseshoe Headrest should be returned to its protective storage case after each use. If the MAYFIELD
Pediatric Radiolucent Horseshoe Headrest is dropped or mishandled, it should be immediately returned to Integra LifeSciences,
Cincinnati, Ohio, for evaluation.
WARNING:
Device must not be used if it has been dropped or mishandled. Doing so may result in serious patient injury.
Time (Min.)
Water Temperature
4:00
Cold water
4:00
Hot water
10:00
60.0°C (140° F)
0:30
Hot water
2:00
82.2°C (180° F)
Detergent and Concentration
N/A
Neutral pH enzymatic (e.g. Endozime® AW Triple Plus with APA,
Ratio 1:128)
Neutral pH detergent
(e.g. Renu-Klenz™ , Steris, Ratio 1:256)
N/A
N/A
11