Dentsply Sirona Midwest Tradition TB Manual Del Operador página 25

Turbina con conexión fija
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Instructions for Use Midwest Tradition / Tradition Pro / Quiet-Air
67 27 155  D3673.201.10.02.02
Spray air pressure in bar (psi)
Spray air flow in Nl/min
Spray water pressure in bar (psi)
Water content in spray ml/min
The indicated pressures can be checked by a service engineer with a
star tester that is placed between the supply hose and the high-speed
handpiece.
3 Technical description
Midwest Tradition / Tradi-
tion Pro / Quiet-Air
2.5 (36)
> 2.5
2.0 ± 0.1 (29)
> 50
23

Publicidad

Tabla de contenido
loading

Tabla de contenido