Please read the product instruction carefully and use the product correctly under the instruction.
1. Dust cap: prevent foreign bodies from entering the product.
2. Lug:used for fixing the fixture.
3. Flat back.
4. Output connector:SPC data output(the bluetooth form can be used for power supply).
5. Stem:used for fixing the fixture.
6. Spindle.
7. Contact point.
8. Battery cap:replace battery.
9. Lcd: display measured values.
10.Lever mounting screw:Used to install lifting bar.
11.Release hole:Used to install lifting release.
12.ON/OFF button:short press to start/shut down.
13.ORIGIN button:Displace the spindle to the position at which the origin is specified, then press
and hold the ORIGIN key for more than three seconds, The origin will be set up, the LCD Will
display"0.000 mm" or "0.00000 in".
14.+/- Button:short-press to switch the sign symbol.
15.mm/in button:short-press to switch between metric and inch measurementmode.
16.ZERO/ABS button:Short press the key at any position to display "INC" and reset to zero at
the same time (entering relative measurement mode). After that, any position can reset at
short press of the key (without affecting absolute origin), and long press the key > 3 seconds
to exit relative measurement mode.
17.PRESET button:Long press the PRESET key for three seconds , The "P" on the LCD will
flash, we now enter the PRESET mode. Long press the PRESET key, the flashing digital
will move from one to the next one, when the digital you want to preset flashing, short press
1. PRODUCT FIGURE AND BUTTONS
ENGLISH
11