Electrical connection of the magnetic stirrer
1.
Connect the cable of the magnetic stirrer to the STIRRER port on
the back of the TitraLab.
The power is supplied via the TitraLab.
Installation of tubes
WA R N I N G
Risk of injury to eyes, skin and respiratory
During the first days of system use, check the screw joint between the syringe
and valve to avoid leaks and the potential escape of reagents.
When working with chemicals and/or solvents, always observe the relevant acci-
dent prevention regulations and wear appropriate personal protective clothing for
eye, face, hand, body and/or respiratory protection.
Make sure that the tubes are never kinked.
Regularly check the system as well as all tubes and connections for
leaks, condition and cleanliness.
To prevent twisting of the tubes, you should always connect flange tube
connections before those with grommets.
The system is delivered with two tubes; these are supplied pre-installed
on the TitraLab and only need to be connected to the relevant titrant
bottle and to the electrode.
Removal of protective stickers (Uninstalled tubes
configuration)
1.
Remove the protective stickers on the burette valve.
2.
Make sure that the pre-installed white O-rings are not removed.
3.
Check that the O-rings are in a horizontal position inside the thread.
English 80
Figure 4 Position of protective stickers
Connection of feed tube
The system is delivered with a tube with bottle closure (DIN 45 thread).
This feed tube is supplied pre-installed on the TitraLab and only needs
to be connected to the titrant bottle.
1.
Connect the tube from the IN socket of the burette valve to the
titrant bottle (NaOH). Screw on the bottle closure to do this.