CAUTION!
Incorrect cleaning could loosen the clamping and cause the drill to acci-
dentally detach during operation.
CAUTION!
After each lubrication, align the head with the drill side downwards.
Operate the head for 30 seconds to remove excess oil. If you notice
any dirt coming out, repeat the entire cleaning and lubrication process.
-Sterilisation
Sterilisation in an autoclave conforming to: EN 13060, EN ISO 17665-1
CAUTION!
- Follow the instructions of the autoclave manufacturer.
- Clean, disinfect and lubricate the handpiece before sterilisation. The
presence of blood residues on the external or internal surfaces can lead
66