Ambient Conditions - Dräger Alcotest 5000 Instrucciones De Uso

Ocultar thumbs Ver también para Alcotest 5000:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

9
Technical data
Measuring principle
Electrochemical sensor

Ambient conditions

During operation
During storage
Measurements
0 mg/L to 0.029 mg/L
0.030 mg/L to 2.5 mg/L
More than 2.51 mg/L
Sensitivity drift
Typically 0.4 % of the measured value per month
Time to re-establish readiness for measurement after a previous
measurement with:
0.025 mg/100 mL 20 s
0.05 mg/100 mL 40 s
0.1 mg/100 mL
>0.15 mg/100 mL 120 s
80 s
–15 °C to +50 °C
(+5 °F to +122 °F)
10 to 100 % rel. hum.
(non-condensing)
600 to 1300 hPa
–20 °C to +60 °C
(–4 °F to +140 °F)
15 to 75 % rel. hum. (10 to 100 % rel.
hum. <48 h)
No alcohol (
Alcohol (
) displayed
↑↑↑
Over range (
0.05 %, BAC
0.10 %, BAC
0.20 %, BAC
>0.30 %, BAC
Technical data
) displayed
) displayed
20 s
40 s
80 s
120 s
en
17

Publicidad

Tabla de contenido
loading

Tabla de contenido