Use
4
Use
4.1

Prerequisites for use

NOTICE
Avoid sampling with a draft. This could falsify the test result.
Observe a wait time of approx. 15 minutes after the last consumption of
alcohol.
The wait time cannot be reduced by rinsing your mouth with water or with
non-alcoholic beverages.
Residual mouth alcohol or aromatic beverages (e.g., fruit juice), mouth
sprays containing alcohol, medication and drops, as well as belching and
vomiting may cause incorrect measuring values.
Rinsing the mouth with water or non-alcoholic drinks does not replace the
wait time.
Draft can falsify the measurement result.
4.1.1
Before sampling
The subject must breathe regularly and calmly before sampling. The subject
must be able to provide the required minimum breath volume for the sampling.
4.1.2
Attaching the wrist strap
The wrist strap should be used to secure the instrument on the wrist.
1.
Guide the fastening lug on the wrist strap through the eye on the instrument.
2.
Pull the top part of the wrist strap through the fastening lug.
4.1.3
Mounting the telescope bar
To reach subjects located further away (e.g., truck drivers), the instrument can
be mounted on a telescope bar with a 1/4" plug-in coupling.
A telescope bar is not included as part of the Dräger Alcotest
Follow the telescope bar manufacturer's instructions for use.
4.1.4
Preparation for use
1.
Ensure that the batteries are charged and inserted correctly.
2.
Ensure that a clean funnel is attached.
3.
Ensure that the instrument is held firmly and securely by using the wrist
strap as intended.