BTL7-A/C/E/G5 _ _ -M _ _ _ _ -K(8)-NEX-SR32/K_ _ _
Micropulse Transducer - Rod Style
6
Calibration procedure (continued)
6.3.2
Adjusting
The detailed procedure for adjusting is
described on page 18 ff.
The factory set null point and end point is replaced by a
new start point and end point and the associated output
values can be adjusted. The start and end values can be
adjusted as you like up to the limits.
Adjustment is possible from serial number
120615000xxxxx xx onwards.
Steps
► Move magnet to the new start position.
► Read new start point by activating the programming
inputs.
► Set the desired start value by activating the program-
ming inputs.
New start value
New null point
Fig. 6-4:
Adjust a new start value
16
english
► Move magnet to the new end position.
► Read new end point by activating the programming
inputs.
► Set the desired end value by activating the program-
ming inputs.
New end value
Fig. 6-5:
Adjust a new end value
Before
6.3.3
Reset
After
Restoring the transducer to its factory settings.
The detailed procedure for the reset is
described on page 20.
Ignition protection
"ec" and "tb"
Before
After
New end point