Breath Sampling With The Funnel; Turning Off The Device; Removing The Mouthpiece; Removing The Funnel - Dräger Alcotest 7000 Instrucciones De Uso

Ocultar thumbs Ver también para Alcotest 7000:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 31
4.6.1.1 Reading the result
The following settings depend on the configuration.
The pre-configured alcohol threshold values determine when
the yellow or red LED illuminates.
Three types of results are possible.
● No alcohol detected (see the fold-out page, Figure 2: D).
In addition, the green LED lights up. The device is ready
for use again after 2 seconds.
Or
● Low level of alcohol detected (see the fold-out page,
Figure 2: E). In addition, the yellow LED lights up. Press
to continue with the next measurement.
Or
● Alcohol detected (see the fold-out page, Figure 2: F). In
addition, the red LED lights up. Press
the next measurement.
Once the recovery time of the sensor (see "Technical data",
page 20) has elapsed, the display indicates that the device is
ready for measurement.
● Repeat step four of the process for taking an automatic
breath sample.
For passive breath samples
Two types of results are possible.
● No alcohol detected (see the fold-out page, Figure 2: G).
In addition, the green LED lights up. Press
with the next measurement.
Or
● Alcohol detected (see the fold-out page, Figure 2: H). In
addition, the red LED lights up. Press
the next measurement.
4.6.2 Taking a manual breath sample (optional)
If the subject cannot provide the required minimum volume of
breath, manual sampling can be activated.
To take a manual breath sample:
1. Follow steps one to four of the process for taking an
automatic breath sample.
2. Briefly press the
button if the circular progress bar is
displayed once breath sampling has begun.
 In the results display, MANUAL is also displayed.
4.6.3 Taking a passive breath sample (optional)
During passive sampling, sampling is manually activated by
the user of the device. Possible applications for this include
testing the ambient air or a person's breath for the presence
of alcohol.
To take a passive breath sample:
1. Hold the ready device without the mouthpiece in the air to
be analyzed.
2. Briefly press the
button to start the measurement.
 If alcohol is detected, the message ALCOHOL is
displayed and the red LED lights up.
 If no alcohol is detected, the message NO ALC. is
displayed and the green LED lights up.
 In the results display, PASSIVE is also displayed.
Instructions for use
|
Dräger Alcotest
to continue with
to continue
to continue with
®
7000
4.7

Breath sampling with the funnel

4.7.1 Taking a breath sample
CAUTION
Hygiene
► Do not touch the upper part of the funnel.
► Always maintain a distance of 2–3 cm from the subject or
object to be tested.
► Do not touch excretions, surfaces or a subject's skin with
the funnel.
► Avoid direct contact with substances to be tested.
► Replace contaminated funnels.
CAUTION
Incorrect measurement results
If multiple samples are taken over a short period of time, there
is a risk of condensation building up inside the funnel and
entering the device.
► Replace the funnel if condensation or drops are visible.
Before taking the breath sample, ensure that the correct
measuring mode has been selected.
1. Remove the funnel from its packaging.
2. Hold the funnel using the grip areas on the sides and
attach it firmly to the top of the device (it can only point in
one direction).
3. Ask the subject to blow evenly and without interruption
into the funnel without touching it.
 A continuous tone signals a sufficient flow of air.
 While the sample is being taken, an animated circular
progress bar that also indicates progress as a
percentage value is displayed (see B and C in Figure 2
on the fold-out page).
 If sampling was successful, the tone ends with a clear,
audible click.
 The display shows a spinning progress bar during an
ongoing analysis process.
 The result is displayed after a few seconds and stored
by the device.
4.7.1.1 Reading the result
For further information see: "Reading the result", page 17.
4.8

Turning off the device

● Press and hold the
seconds.
The device turns off automatically after 4 minutes of not
being used.
4.9

Removing the mouthpiece

● Push the mouthpiece out of the mouthpiece receptacle in
an upward motion and dispose of it (see Figure 1 on the
fold-out page).

4.10 Removing the funnel

● Hold the funnel using the grip areas on its sides, pull it out
of the device in an upward motion and dispose of it.
Use
button. The device turns off after 3
|
en
17