Sterilization
After performing the cleaning instructions specified above, perform one of the
following sterilization cycles.
Ethylene Oxide (EtO)
•
Double wrap camera head and cable prior to sterilization.
Preconditioning parameters
Temperature
Chamber Humidity
Vacuum Set Points
Time
Exposure
Concentration (100% EtO)
Temperature
Time
Chamber Humidity
Aeration parameters
Aeration Time
Temperature
Steris® System 1
1. Clean and prepare the urology camera head and cable as recommended in
the Cleaning section.
2. Sterilize the urology camera head and cable using Steris® System 1 with
Steris® Sterilant 20.
3. Allow the urology camera head, cable, and scope to completely dry before
reassembly. Any moisture may cause the camera window to fog during use.
Sterrad®
1. Clean and prepare the urology camera head and cable as recommended
in the Cleaning section.
2. Sterilize the urology camera head and cable using the Sterrad™ NX or
100S Sterilization System.
Storage
Never store the device in a non-ventilated, humid environment such
as a carrying case. This may present an infection control risk.
55°C (131°F)
70% RH
1.3 psia
30 minutes
725 mg/L
55 ± 2°C (131 ± 5°F)
1 hour
70 ± 5% RH
12 hours
35 – 54°C (95 – 129°F)
19