Sterilization
After performing the cleaning instructions specified above, perform one of the
following sterilization cycles.
Ethylene Oxide (EtO)
Double wrap camera head and cable prior to sterilization.
Preconditioning parameters
Temperature
Chamber Humidity
Vacuum Set Points
Time
Exposure
Concentration (100% EtO)
Temperature
Time
Chamber Humidity
Aeration parameters
Aeration Time
Temperature
Steris® System 1
1. Clean and prepare the urology camera head and cable as recommended in
the Cleaning section.
2. Sterilize the urology camera head and cable using Steris® System 1 with
Steris® Sterilant 20.
3. Allow the urology camera head, cable, and scope to completely dry before
reassembly. Any moisture may cause the camera window to fog during use.
Sterrad®
1. Clean and prepare the urology camera head and cable as recommended
in the Cleaning section.
2. Sterilize the urology camera head and cable using the Sterrad™ NX or
100S Sterilization System.
Storage
Never store the device in a non-ventilated, humid environment such
as a carrying case. This may present an infection control risk.
55°C (131°F)
70% RH
1.3 psia
30 minutes
725 mg/L
55 ± 2°C (131 ± 5°F)
1 hour
70 ± 5% RH
12 hours
35 – 54°C (95 – 129°F)