Problem
After the device is started, the
messages "Maximum number of
scans and color measurements has
been exceeded. The oldest scan and
color measurements are overwritten"
appear or "The number of scans and
color measurements is near the limit.
Please delete measurements which
are no longer needed" is displayed in
red on the background of the main
screen.
An error occurred during the
measurement. Please repeat the
measurement.
Warning: Some system components
were not found!
No free method number available!
Measurement cannot be performed
(division by 0)! Please repeat the
measurement!
Barcode already linked in another
method.
MACHEREY‑NAGEL GmbH & Co. KG
If you still have questions after reading the instructions or need technical assistance, contact:
MACHEREY‑NAGEL GmbH & Co. KG
Valencienner Str. 11 · 52355 Düren · Germany
Telephone: +49 2421 969‑0
Email: info@mn‑net.com
www.mn‑net.com
This product can be sold in the UK. The conformity is declared by the UKCA logo on
the device type label and the product label. A separate declaration of conformity can be
requested from MACHEREY‑NAGEL or your responsible distributor.
Contact of our UK authorized representative:
COMPLY EXPRESS LIMITED
Unit C2 Coalport House, Stafford Court, Stafford Park 1, Telford
TF3 3BD, UK.
Spectrophotometer NANOCOLOR
Quickstart Guide NANOCOLOR
Cause
The memory for scan
and color measurements
(100 results) is full or
nearly full.
The measurement was
interrupted.
There is a problem with
the internal USB hub.
The maximum possible
number of special
methods has been
created.
Division by zero
occurred during
the measured value
calculation.
The linked barcode is
already linked in another
special method.
/
II and VIS II
® UV
VIS
® UV
Solution
Depending on the message, delete the
measurement data from the measured value
memory. If the message is ignored, the oldest
data set will be overwritten if the memory
is full. Please note that during automatic
overwriting, color references can also be
overwritten.
Wait until the measurement has been
completed in full before you remove the
cuvette.
Contact MACHEREY‑NAGEL
Remove special methods which are no longer
needed in order to be able to create a new
method.
Ensure that all cuvettes inserted were
positioned according to the instructions for the
corresponding method.
Delete the barcode link first in the other
method before the barcode is to be linked in
the new method.
/
II / VIS II V 2.00/09‑2022
VIS
23