CAUTION
Risk of burns due to high leakage currents
Simultaneous contact with two metallic parts can result in high leakage currents, which
can cause skin burns.
z Do not touch any metallic parts during use.
Electric shock due to water entering the device
Water ingress may cause a short circuit in the device and endanger the user.
z Disconnect live parts from the power supply before cleaning them.
z The electrical contacts on the B.Box Evo Rechargeable Battery (B.Box Evo Battery
Stand) must not come into contact with liquids.
Crushing injuries due to components with magnets
When the B.Grip Evo holder module is placed in close proximity to the B.Spot Evo, B.Light
Clear Evo, and B.Light Restore Evo application modules, skin pinching may occur due to the
resulting magnetic attraction. Skin bruising may also occur when attaching the B.Box Evo
to the B.Box Evo Stand or the B.Box Evo Battery Stand.
z Follow the instructions on the corresponding components and do not reach
between the holder module and the application modules or the B.Box Evo and the
B.Box Evo Stand or the B.Box Evo Battery Stand.
Increase in vital signs due to incorrect operation of the device
User errors may result in elevated vital signs, e.g., higher blood pressure, especially
among untrained users.
z When purchasing the BEMER Therapy System Evo, ensure that you are given an
introduction to the system by an official and certified BEMER partner.
z Always read and familiarize yourself with the instructions for use included in the
scope of delivery.
z Observe the safety notes.
18
CAUTION
Allergic reactions due to material incompatibility
The materials used in the applicators may cause skin reactions due to intolerances.
z If this occurs, discontinue use of the therapy system and consult your doctor.
NOTE
Increased surface temperature of the LLLT application modules
due to high ambient temperature
At ambient temperatures above 35°C, the surface of the light therapy application
modules can reach 44°C.
z If the ambient temperature exceeds 35°C, allow the light therapy application
modules to cool for at least 10 minutes between treatments.
Performance degradation of portable RF communication devices
due to electromagnetic forces
The performance of the respective devices may be affected by electromagnetic forces.
z Do not use portable RF communication devices (including accessories such as
antenna cables and external antennas) within 12 inches (30 cm) of any part of the
BEMER Therapy System Evo, including the cables specified by the manufacturer.
Using the device in an unsuitable environment
Use of this device in the immediate vicinity of other equipment and / or in a humid
environment can lead to malfunctions.
z Do not use this device when it is directly adjacent to or stacked on/under other
equipment.
z If such use is unavoidable, monitor the device as well as the other equipment to
ensure normal operation.
z Only use the BEMER Therapy System Evo in dry rooms.