DRY FLOATATION Air Insert: Cleaning and Disinfecting
Warnings:
- Inflation valve(s) MUST be closed to make sure that water does not enter the product during cleaning or disinfecting.
- Follow all manufacturer bleach container and germicidal disinfectant safety guidelines.
- Note about sterilization: High temperatures accelerate aging and will damage the product assembly. The products featured in this
manual are not packaged in a sterile condition, nor are they intended or required to be sterilized prior to use. If institution protocol
demands sterilization: First, follow the cleaning and disinfecting instructions; then, open the inflation valve and use the lowest pos-
sible sterilizing temperature, but not to exceed 70°C (158°F), for the shortest time possible. DO NOT autoclave.
Cautions: DO NOT machine wash or machine dry the air insert. Use of the following may cause damage to the cushion: abrasives
(steel wool, scouring pads); caustic, automatic dishwasher detergents; cleaning products that contain petroleum or organic solvents,
including acetone, toluene, Methyl Ethyl Ketone (MEK), naphtha, dry cleaning fluid, adhesive remover; oil-based lotions, lanolin, or phenolic-
based disinfectants; ozone gas; ultraviolet light cleaning methods.
To Clean the Air Insert: Remove the air insert from the cover and foam pad, close valve(s) and place in a
large sink. Hand wash, using a mild, liquid hand soap, hand dishwashing liquid, laundry detergent, or other
multipurpose detergent (Follow product label instructions.) Gently scrub all air insert surfaces, using a soft plastic
bristle brush, a sponge or a cloth. Rinse thoroughly with clean water. Air dry. DO NOT expose to direct sunlight.
Caution: Rinse thoroughly. Detergent residue may cause air cells to stick together.
To Disinfect the Air Insert: Repeat the cleaning instructions, using 1 part household liquid bleach per 9 parts
water. Keep the product wet with the bleach solution for 10 minutes. Rinse thoroughly with clean water. Air dry.
Limited Warranty
Limited Warranty Term, from the date the product was originally purchased: DRY FLOATATION Air Insert: 36 months; Cover and Foam Pad: 18
months. See also the Limited Warranty supplement provided with your product, or contact Customer Support.
Contact Information
ROHO, Inc.
Administrative Offices: 100 North Florida Avenue, Belleville, IL 62221-5429, U.S.A.
Mailing/Shipping Address: 1501 South 74th Street, Belleville, IL 62223-5900, U.S.A.
U.S.A.: 800-851-3449 • Fax: 888-551-3449 • Customer Support: orders.roho@permobil.com
Outside the U.S.A.: 618-277-9150 • Fax: 618-277-6518 • intlorders.roho@permobil.com
permobilus.com
Products may be covered by one or more U.S. and foreign patents and trademarks, including AGILITY
Product information is changed as needed; current product information is available at permobilus.com.
ROHO, Inc. is part of Permobil Seating and Positioning.
Rev. 2018-10-19
© 2013, 2018 Permobil
T20203
Custom ROHO AGILITY Back Cushion Supplement
2
+
1:9
DRY FLOATATION and ROHO
.
®
®