9. Always check that the device is in proper working condition before
use. Always check that the device is in proper working condition be-
fore use. If device malfunctions are indicated (e.g. sluggish pistons,
leaks), stop pipetting immediately and refer to the section "Trou-
bleshooting"; see Troubleshooting, p. 56 . Contact the manufac-
turer, if necessary.
2.2 Purpose
This is an air displacement pipette for pipetting aqueous solutions of
medium density and low to medium viscosity.
2.3 Limitations of use
This instrument is intended for pipetting samples, within the following
limitations:
• Operating temperature of instrument and reagent should be be-
tween +15 °C and +40 °C (59 °F to 104 °F) (other temperatures upon
request)
• Vapor pressure up to 500 mbar
• Viscosity: 260 mPa s 260 mPa s
For viscous media, the speed must be adjusted if necessary.
2.4 Application restrictions
Viscous and wetting liquids may compromise volumetric accuracy. Volu-
metric accuracy may also be affected when pipetting liquids whose tem-
perature deviates from the ambient temperature by more than ± 1 °C/±
1.8 °F.
2.5 Operating exclusions
The user is responsible for checking the compatibility of the device with
the intended application. The device cannot be used:
Transferpette® S 997207
Operating manual
39