Descargar Imprimir esta página

WIKA MH-4-CAN Manual De Instrucciones página 7

Ocultar thumbs Ver también para MH-4-CAN:

Publicidad

Idiomas disponibles

Idiomas disponibles

4. Mounting
4. Mounting
4.1 Mechanical mounting
Only use the pressure transmitter if it is in perfect condition with respect to safety.
Prior to commissioning, the pressure transmitter must be subjected to a visual inspection.
Leaking liquid is indicative of damage.
Obvious damage must be reported immediately.
Mounting the instrument
The max. torque depends on the mounting point (e.g. material and shape). If you have any questions,
please contact our application consultant.
→ For contact details, see chapter 1 "General information" or the back page of the operating instructions.
1.
Seal the sealing faces.
2
At the mounting point, screw the pressure sensor in hand-tight.
.
3
Tighten with a torque wrench using the spanner flats.
.
Hexagon at the case, spanner width 22 mm:
- Open-ended spanner: 30 Nm ±5 %
- Socket wrench: 50 Nm ±5 %
Optional hexagon at the process connection, spanner width 27 mm:
- Open-ended spanner: 50 Nm ±5 %
- Socket wrench: 50 Nm ±5 %
Operation with vented electrical output
WIKA operating instructions pressure sensor, model MH-4-CAN
Hexagon (SW 22) integrated
into case
22
Additional hexagon (SW 27)
above the process connection
27
EN
7

Publicidad

loading