Repair and maintenance of the pipettor
Cleaning:
External surfaces of the pipetting pushbutton, the ejector
pushbutton, the handgrip, the shaft nut and the adjust-
ment knob may be cleaned using a cloth dampened in
isopropyl alcohol. Remaining parts removed from the
pipettor during pipettor disassembly may be washed
with distilled water or isopropyl alcohol.
Sterilization:
The pipettors can be sterilized in the autoclave at 121°C
for 20 minutes. After sterilization, the pipettor should be
dried and cooled down to room temperature.
It is recommended:
- to sterilize the pipettors in autoclave with an initial vacuum
and drying cycle.
- prior to sterilization to unscrew the shaft nut slightly in
the VWR2-VWR1000 pipettors, and unscrew the shaft
slightly in the VWR5000 and VWR10000. After autoclaving
these parts should be screwed tight again.
- to test pipettor calibration every 10 sterilization cycles.
Note: VWR pipettors should be autoclaved with the
filter removed (VWR5000 and VWR10000).
After autoclaving, check that the shaft nut is tightly
fitted to the pipettor handle.
20
User replaceable accessories and spare parts
Description
Tip, 10 μl
Tip, 200 μl
Tip, 1000 μl
Tip, 5000 μl
Tip, 10000 μl
A: Pushbutton*
A1: Pushbutton A2: Knob
B: Adjustment knob
C: Shaft
D: Ejector
F: Shaft nut
G: Piston assembly
I: O-ring
J: Seal
K: Calibration key
L: Filter
M: Tip ejector cap
N: Ejector pushbutton
* Please ask your VWR office for spare parts.
Technical service
Web Resources
Visit the VWR website at www.vwr.com for:
• Complete technical service contact information
• Access to the VWR Online Catalogue, and information
about accessories and related products
• Additional product information and special offers
Contact us
For information or technical assistance contact your local
VWR representative or visit. www.vwr.com.
Version: 1
Issued 17/03/2014
ENGLISH
Cat. No.
613-0334
613-0241
613-0273
613-0830
613-0831
21