4.3 Care of the connection tube
Dry the connection tube after every inhalation session:
• Connect the connection tube to the compressor.
• Switch your compressor on.
• Leave the compressor to work until the air flowing through
the tube has removed all traces of any condensation in the
tube.
4.4 Replacing the air filter
Take the air filter out:
• Pull the filter holder out of the
compressor. Use a small
screwdriver, for example, to
prise the filter holder carefully
out of the compressor.
Replace the air filter:
• Pull the old air filter out of the filter holder and fit the new fil-
ter in its place.
NOTE! Only use air filters provided by PARI GmbH for use
with your compressor. If air filters that are not designed for
the compressor are used, the compressor may be
damaged.
• Insert the filter holder back in the compressor.
4.5 Inspecting
Inspect all product components after each cleaning and disin-
fection. Replace any broken, misshapen or seriously dis-
coloured parts.
– 34 –