4.2 Multiple patient use
Cleaning and disinfection cycles
Nebuliser with connection
tubing, mask and air filter
Compressor housing
Nebuliser and mask
The nebuliser and mask are not suitable for use by multiple
patients. Use a separate nebuliser (including connection
tubing) and a separate mask for each patient.
Air filter
All air filters must only be used for one patient. Replace the air
filter before every patient change [see: Replacing the air filter,
page 34].
Disinfecting the compressor housing
For disinfection, use a standard, alcohol-based disinfectant
(e.g., isopropanol). Regarding the use and quantity of the disin-
fectant, it is imperative to follow the information for use of the
agent used.
To disinfect the compressor, proceed as follows:
• If the compressor is visibly dirty, clean the compressor be-
fore disinfecting it [see: Cleaning the compressor housing,
page 32].
• Moisten a cloth with the disinfectant.
NOTE! Danger of device fault due to liquid penetration.
Never spray any liquids onto the compressor or the power
cord. If liquid does get into the compressor, it must not be
used under any circumstances. Before starting the com-
pressor again, contact the Technical Service department or
the Service Center of PARI GmbH.
• Wipe the outer surfaces of the housing off thoroughly with
the cloth.
– 33 –
Replace before each patient change
Disinfect before each patient change