Cleaning The Instrument; Storage; Shipment - MSA Altair 5X Manual De Instrucciones

Ocultar thumbs Ver también para Altair 5X:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Use of the fibrous dust filter or the incorrect paper filter for
the measurement of reactive gasses could cause
erroneous readings.
Place the new paper-like filter into the recess in the back of the
instrument. If it is to be used, place the fibrous dust filter into the
clear filter cover.
5. Replace the O-ring in the recess.
6. Re-install the clear filter cover on the back of the instrument.

4.6 Cleaning the Instrument

Clean the exterior of the instrument regularly using only a damp cloth.
Do not use cleaning agents as many contain silicones which damage
the combustible sensor.

4.7 Storage

When not in use, store the instrument in a safe, dry place between
65°F and 86°F (18 °C and 30 °C). After storage, always recheck
instrument calibration before use. If not to be used in 30 days,
remove battery pack or connect it to a charger.

4.8. Shipment

Pack the instrument in its original shipping container with suitable
padding. If the original container is unavailable, an equivalent container
may be substituted.
! CAUTION
4-7

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Altair 5x ir

Tabla de contenido