Valores De Medición - Vega VEGAPULS C 23 Instrucciones De Servicio

Sensor de radar para la medición continua de nivel
Ocultar thumbs Ver también para VEGAPULS C 23:
Tabla de contenido

Publicidad

14 Anexo
Register Num-
Register Name
ber
3601
Application liquid
3602
Application solid
14.5 Valores de medición
Register Number
Register Name
100
Status
104
PV Unit
106
PV
108
SV Unit
110
SV
112
TV Unit
114
TV
116
QV Unit
118
QV
54
Data type
Configurable Values
enum16
0 =Storage tank, 1=
Storage tank (product agi-
r/w
tation), 2=Cargo tank,
3=Reactor tank, 4=Do-
sing tank, 5=Stilling pipe,
6=Bypass, 7=Outside
of plastic tank, 8=Outsi-
de of mobile plastic tank,
9=Level of open water,
10=Open flume, 11=Water
weir, 12=Demonstration,
13=Pump station, 14=Co-
llection tank
enum16
0=Silo, 1= Bunker Big,
2=Bunker fast filling,
r/w
3=Profil registration heap,
4=Breaker mill, 5=De-
monstration
Data type
Note
Access
enum8
Bit 0: Invalid Measurement Value PV, Bit 1: Invalid Measu-
rement Value SV, Bit 2: Invalid Measurement Value TV, Bit
ro
3: Invalid Measurement Value QV
enum16
32=Degree Celsius, 33=Degree Fahrenheit, 40=US Ga-
llons, 41=Liters, 42=Imperial Gallons, 43=Cubic Meters,
ro
44=Feet, 45=Meters, 46=Barrels, 47=Inches, 48=Cen-
timeters, 49=Millimeters, 111=Cubic Yards, 112=Cubic
Feet, 113=Cubic Inches
float32
Primary Variable in Byte Order CDAB
ro
enum16
Unit Code
ro
float32
Secondary Variable in Byte Order CDAB
ro
enum16
Unit Code
ro
float32
Third Variable in Byte Order CDAB
ro
enum16
Unit Code
ro
float32
Quarternary Variable in Byte Order CDAB
ro
VEGAPULS C 23 • Protocolo Modbus y Levelmaster
Unit
Default Va-
lue
-
Storage
tank
-
Silo

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Vegapuls c 11

Tabla de contenido