12
The following set point types can be selected:
– High (High value has to be set)
– Low (Low value has to be set)
– Between (High and Low value has to be set)
– Outside (High and Low value has to be set)
For Conductivity, the following additional options may be selected:
– USP (% safety margin below U.S. Pharmacopoeia limits)
– EP PW (% safety margin below European Pharmacopeia
limits for Purified Water)
– EPWFI (%safety margin below European Pharmacopeia
limits forWater for Injection)
– JP Cond (% safety margin below Japanese Pharmacopoeia limits)
After setting the set point value(s) a Relay can be selected from the relays available
on the transmitter, and configured for that set point. The Relay delay is set to 10
seconds and the Hysteresis is set to 5%. These parameters can be configured from
the Relay menu.