Overview
Nameplate
The nameplate contains the most important data for
identification and use of the sensor.
Type
DulcoLevel
El. connection
18 - 30 V
1.8 W
0.1 A
max
Measuring range
0.1 - 5m
IP66
Notice: You need the last 4 digits of the serial number
(Ser.No. on the nameplate) to connect the sensor with
the DULCONNEX Blue App (see Initial Start-up).
In the example shown, this corresponds to the digits
"8234".
Notice: The QR code on the nameplate is for
ProMinent internal purposes only.
The device meets the requirements for electrical safety
according to EN 61010-1. The device meets the require-
ments of the Radio Equipment Directive 2014/53/EU and
the RoHS Directive 2011/65/EU and is CE marked.
ProMinent GmbH
Im Schumachergewann 5-11
69123 Heidelberg, Germany
Mat.No. 1120573
Ser.No.
e. g. 20220808234
MD:
e. g. 07/2022
Contains FCC ID: QOQGM210P
Contains IC ID: 5123A-GM210P
Function
The level sensor DULCOLEVEL is a radar sensor for
level measurement of all types of liquids in plastic
tanks and IBC storage tanks:
Measures the distance between the sensor and
the surface of the liquid.
Contactless measurement through the plastic wall
of the tank.
Suitable for all ProMinent plastic storage tanks as
well as plastic tanks from other suppliers.
Configure the sensor by using the DULCONNEX
Blue App.
Notice: Use of a Mixer
When using a mixer, the height of the liquid level may
change. This can lead to incorrect measurement
results. An even liquid level is required for a reliable
measurement result. Select the measuring point out-
side the flow if possible.
7
EN