• The message "----" indicates that the calibration is complete. Remove the weight
from the platform.
• Switch off the unit using the "ON/OFF" button
C. SBS-PT-60SNL; SBS-PT-150SNL:
• While holding the "ZERO" and "TARE" buttons simultaneously, turn on the unit
with the "I/O" switch. The display will show "CAL".
• Press the TARE button. The display will indicate the calibration value set in the
Setup Menu.
• Place the weight with the displayed weight on the weighing platform.
• After the reading has stabilized, press the "TARE" button.
• The message "----" indicates that the calibration is complete. Remove the weight
from the platform and turn off the unit by setting the switch to the "O" position.
D. SBS-PT-30SN; SBS-PT-150SN:
• While holding the "UNIT" and "TARE" button, turn on the unit with the "ON/OFF"
button. The display will show "CAL".
• Press the "UNIT" button again. The display will indicate the calibration value.
• Place the weight with the displayed weight on the weighing platform.
• After the reading has stabilized, press the "UNIT" button.
• The display of a string of numbers (the number of measurements taken so far)
indicates the completion of the calibration. Remove the weight from the platform.
• Switch off the unit using the "ON/OFF" button
E. SBS-LW-40:
• When the scale is off, hold the "UNIT" button for a few seconds, then press the
"ON/OFF" button. The display will show "C=xx".
• Press the "UNIT" button to select the calibration unit (kg; lb).
• Press the "ZERO/TARE" button. The display will indicate "CAL", followed by the
number of measurements taken so far.
• Press the UNIT button. The display will show "-----" followed by the calibration
value.
• Place a weight of the indicated weight on the platform.
• Press the UNIT button.
• The message "-----" and the subsequent display of the number of measurements
taken so far indicate the completion of the calibration process.
• Remove the weight from the platform.
• Switch off the unit using the "ON/OFF" button
CAUTION! (applies to all models)

Publicidad

loading