EN – ENGLISH
4. Follow the instructions listed below and set washer
machine to these exact parameters:
Phase
Time
Water
(Min.)
Temperature
Pre-wash 1
4:00
Cold water
Enzyme
4:00
Hot water
Wash
Wash 1
10:00
60.0ºC
(140º F)
Rinse 1
0:30
Hot water
Thermal
2:00
82.2ºC
Rinse**
(180º F)
** Optional phase for disinfection of components –
minimum water temperature as indicated or per worker
manufacturer specifications for the thermal rinse cycle.
NOTE – Any deviation from this guideline could result in
damage to the equipment as well as improper cleaning
results.
Rinse with purified water. Do not perform if parameters
cannot be achieved.
5. Remove from washer and dry completely if needed.
6. Inspect equipment to make sure there is no visible
organic debris or residue from cleaning agent.
Repeat process if any visible soil is detected.
The life expectancy of the MAYFIELD products is
expected to be as long as 7 years.
Maintenance and Care
To ensure proper function and to extend the life and
performance of the equipment, Integra LifeSciences
recommends the following:
Recommended Action
Return the device to the Integra
LifeSciences Repairs department for
detailed inspection and servicing.
Request that Integra NeuroSpecialists
perform routine inspections of the
device
Detergent and
Concentration
N/A
Neutral pH
enzymatic (e.g.
Endozime® AW
Triple Plus with APA,
Ratio 1:128)
Neutral pH
detergent
(e.g. Renu-Klenz™ ,
Steris, Ratio 1:256)
N/A
N/A
Recommended
Frequency
Once / year
Twice / year
8
In the absence of proper care and servicing of the device,
negative effects may be seen after repeated processing
over time which may lead to reduced performance.
Contact information: See the Service and Repair section
for contact information on how to return your device for
periodic servicing and to request periodic inspections.
See Inspection and/or Service notes section for routine
checks to be performed on the device.
NOTE – Any serious incident that has occurred in relation
to the device for the user and/or the patient should be
reported to the manufacturer and the competent authority
of the member state in which the user and/or patient is
established.
Device Disposal
NOTE: Follow hospital procedures for disposal of this device.
Warranty
INTEGRA LIFESCIENCES CORPORATION ("INTEGRA")
warrants to the original purchaser only that each new
MAYFIELD product is free from manufacturing defects
in material and workmanship under normal use and
service for a period of one year (except as otherwise
expressly provided as to accessory items) from the date
of delivery by INTEGRA to the first purchaser, but in no
event beyond the expiration date stated on any product
labeling.
• Surgical instruments are guaranteed to be free
from defects in material and workmanship when
maintained and cleaned properly and used normally
for their intended purpose.
• Any covered product that is placed by INTEGRA
under a lease, rental or installment purchase
agreement and that requires repair service during
the term of such placement agreement shall be
repaired in accordance with the terms of such
agreement.
If any covered defect occurs during the warranty period
or term of such placement agreement, the purchaser
should communicate directly with INTEGRA's home
office. If purchaser seeks to invoke the terms of this
warranty, the product must be returned to INTEGRA at its
home office. The defective product should be returned
promptly, properly packaged and postage prepaid. Loss
or damage in return shipment to INTEGRA shall be at
CUSTOMER's risk. INTEGRA's sole responsibility under
this warranty shall be repair or replacement, at INTEGRA's
sole discretion at INTEGRA's expense, subject to the
terms of this warranty and applicable agreements.