Exit The Calibration Procedure; Reset Calibration Defaults; Maintenance - Hach LDO Manual Del Usuario

Ocultar thumbs Ver también para LDO:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 37
Option
Description
Hold
The sensor output value is held at the current measured value during the calibration procedure.
Transfer A preset value is sent during calibration. Refer to the controller user manual to change the preset
value.
5. The controller will show:
• "Press ENTER when stabilized"
• The current dissolved oxygen measurement
• The current temperature measurement
6. When the measurement is stable, push ENTER. The display will show an entry screen.
Note: The measurement will usually stabilize in 2 to 3 minutes.
If the value does not stabilize, the display will show "Unable to Calibrate" followed by an error
message.
Table 4
shows the error message and resolution for calibration problems.
Message
Description
Cal fail, offset high The calculated offset value is too high.
Cal fail, offset low
The calculated offset value is too low.
Cal fail, unstable
The value did not stabilize in the maximum allowed calibration time.

Exit the calibration procedure

1. During calibration, push the BACK key. Three options are shown:
Option
Description
ABORT
Stop the calibration. A new calibration must start from the beginning.
BACK TO CAL Return to the current calibration.
LEAVE
Exit the calibration temporarily. Access to other menus is allowed while the calibration
continues in the background. A calibration for a second sensor (if present) can be started.
To return to the calibration, push the MENU key and select Sensor Setup, [Select Sensor].
2. Select one of the options. Confirm.

Reset calibration defaults

Calibration settings can be reset to the factory defaults. Gain and offset values are set to 1.0 and 0.0,
respectively.
1. Go to MENU>SENSOR SETUP>[Select Sensor]>CALIBRATE>RESET CAL DEFLT.
2. The display will show a confirmation message. Confirm to reset the sensor to the factory default
calibration curve.

Maintenance

Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
14 English
Table 4 Sample cal error messages
D A N G E R
Resolution
Repeat the calibration.
Repeat the calibration.
Repeat the calibration.