Descargar Imprimir esta página

Dräger X-act 7000 Manual Tecnico página 48

Ocultar thumbs Ver también para X-act 7000:

Publicidad

Idiomas disponibles

Idiomas disponibles

|
en
Operation
Type
Approved batteries:
Alkaline batteries (Zn-MnO
– Panasonic AA LR6XWA LR6
– Duracell Plus Power MN 1500
– VARTA 4706 Longlife Max Power AALR6 LR6
– Industrial by Duracell AA (LR6) ID1500
Lithium batteries (FeS
– Energizer Ultimate Lithium AA
– Ansmann Industrial Lithium AA
The battery status indicator of the X-act 7000 is designed for alkaline batteries with
a rated voltage U
voltage U
4.2
Preparation for operation
Explosion hazard!
The use of accessories that are not approved for use in explosion-hazard areas
may lead to the ignition of gas-charged atmospheres.
► Only use accessories that have been approved by Dräger for use in explosion-
page 47.
Prior to commissioning, insert the batteries outside of the explosion-hazard area
(Graphic A/13) and attach the carrying strap (see Graphic A/14). Mount the
approved components for probe measurement.
4.2.1
Insert the batteries
1. Unscrew four torx screws on the battery cover (Graphic A/9) on the rear of the
2. Insert the batteries included in the scope of delivery (Graphic A/13). Pay
3. Close the battery cover again using the included tool. Make sure the battery
4.2.2
Attach the carrying strap
1. Attach the carrying strap included to the eye on the rear of the device (Graphic
48
, U
2
of 1.5 V. The use of the specified lithium batteries with a rated
N
of 1.5 V is permissible, but may lead to a false battery status reading.
N
WARNING
hazard areas. Further information available on the product page at
www.draeger.com.
Please observe the battery instructions. For further information see: "Batteries",
device using the included tool.
attention to polarity.
cover has been sealed shut.
A/14). Make sure the press studs are on the inside. This prevents the belt from
opening unintentionally.
, U
= 1.5 V)
2
N
= 1.5 V)
N
Technical Manual
|
®
X-act
7000

Publicidad