Warnings
Remote pump - battery
low
Languages - not found
Languages - unknown
Languages - corrupted
Substances - not found
Substances - corrupted
Databases - incompatible The version of the sub-
Locations - corrupted
Memory - exceeded
|
Technical Manual
®
X-act
7000
Cause
The charge status of the
rechargeable battery for
the X-am pump is low
The language table could
not be found in the mem-
ory
The version of the lan-
guage table is incompati-
ble with the X-act 7000
firmware
The language table is
faulty
The substance table could
not be found in the mem-
ory
Substance table is faulty
stance table is incompati-
ble with the X-act 7000
firmware
The measurement location
table is faulty
The memory of the X-act
7000 is almost exhausted
Troubleshooting
Remedy
Promptly recharge the X-
am pump
The X-act 7000 can still be
used, but only in English;
install the current lan-
guage table on the X-act
7000 with CC-Vision
The X-act 7000 can still be
used, but only in English;
install the current lan-
guage table on the X-act
7000 with CC-Vision
The X-act 7000 can still be
used, but only in English
The X-act 7000 can still be
used, but "Unknown gas"
is always displayed as the
gas name; install the cur-
rent substance table on
the X-act 7000 with CC-
Vision
The X-act 7000 can still be
used, but "Unknown gas"
is always displayed as the
gas name; install the cur-
rent substance table on
the device with CC-Vision
The X-act 7000 can still be
used, but "Unknown gas"
is always displayed as the
gas name; install the cur-
rent substance table on
the X-act 7000 with CC-
Vision
Reconfigure the measure-
ment locations with CC-
Vision
If possible, remove unnec-
essary measurement loca-
tions from the
measurement location
table with CC-Vision
|
en
57