Dismounting, Return And Disposal - WIKA 116.15 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

8. Dismounting, return and disposal

8. Dismounting, return and disposal
Personnel: Skilled personnel
Protective equipment: Safety goggles, protective gloves
Tool: Open-ended spanner
WARNING!
EN
Physical injuries and damage to property and the environment
through residual media
Upon contact with hazardous media (e.g. oxygen, acetylene, flammable or
toxic substances), harmful media (e.g. corrosive, toxic, carcinogenic, radioac-
tive), and also with refrigeration plants and compressors, there is a danger of
physical injuries and damage to property and the environment.
WARNING!
Risk of burns
During dismounting there is a risk of dangerously hot media escaping.
WARNING!
Physical injury
When dismounting, there is a danger from aggressive media and high
pressures.
8.1 Dismounting
1. Before dismounting, make sure that the measuring point is not pressurised and has
cooled down.
2. Dismount the pressure gauge with an open-ended spanner using the spanner flats of
the process connection
14
Before storage of the dismounted instrument (following use) wash or
clean it, in order to protect persons and the environment from exposure to
residual media.
Wear the requisite protective equipment (see chapter 3.6 "Personal
protective equipment").
Let the instrument cool down sufficiently before dismounting it!
Wear the requisite protective equipment (see chapter 3.6 "Personal
protective equipment").
Observe the information in the material safety data sheet for the corre-
sponding medium.
Disconnect the pressure gauge once the system has been depressurised.
WIKA operating instructions models 116.15 and 116.18

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

116.18

Tabla de contenido