Transport, Packaging And Storage; Transport; Packaging And Storage; Commissioning, Operation - WIKA CTD9100-ZERO Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

4. Transport, packaging and storage / 5. Commissioning, operation

4. Transport, packaging and storage

4.1 Transport

Check the model CTD9100-ZERO temperature dry-well
calibrator for any damage that may have been caused by
transport.
Obvious damage must be reported immediately.
CAUTION!
Damage through improper transport
With improper transport, a high level of damage
to property can occur.
When unloading packed goods upon delivery
as well as during internal transport, proceed
carefully and observe the symbols on the
packaging.
With internal transport, observe the
instructions in chapter 4.2 "Packaging and
storage".
If the instrument is transported from a cold into a warm
environment, the formation of condensation may result
in instrument malfunction. Before putting it back into
operation, wait for the instrument temperature and the room
temperature to equalise.

5. Commissioning, operation

Personnel: skilled personnel
Protective equipment: safety goggles, protective gloves
Only use original parts (see chapter 11 "Accessories").
WARNING!
Physical injuries and damage to property
and the environment caused by hazardous
media
Upon contact with hazardous media (e.g.
oxygen, acetylene, flammable or toxic
substances), harmful media (e.g. corrosive,
toxic, carcinogenic, radioactive), and also with
refrigeration plants and compressors, there is
a danger of physical injuries and damage to
property and the environment.
For these media, in addition to all standard
regulations, the appropriate existing codes or
regulations must also be followed.
Wear the requisite protective equipment (see
chapter 2.6 "Personal protective equipment").
WIKA operating instructions temperature dry well calibrator, model CTD9100-ZERO

4.2 Packaging and storage

Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection
during transport (e.g. change in installation site, sending for
repair).
Permissible conditions at the place of storage:
Storage temperature: 0 ... 60 °C (32 ... 140 °F)
Humidity: 35 ... 85 % relative humidity (no condensation)
Avoid exposure to the following factors:
Direct sunlight or proximity to hot objects
Mechanical vibration, mechanical shock (putting it down
hard)
Soot, vapour, dust and corrosive gases
Hazardous environments, flammable atmospheres
Store the temperature dry-well calibrator in its original
packaging in a location that fulfils the conditions listed above.
If the original packaging is not available, pack and store the
instrument as described below:
1. Place the instrument, along with shock-absorbent
material, in the packaging.
2. If stored for a prolonged period of time (more than 30
days), place a bag containing a desiccant inside the
packaging.

5.1 Operating position

DANGER!
Danger to life caused by electric current
Upon contact with live parts, there is a direct
danger to life.
Operation using a defective power cord (e.g.
short-circuit from the mains voltage to the
output voltage) can result in life-threatening
voltages at the instrument!
The operating position of the temperature dry-well
calibrator is in the vertical orientation, since this
guarantees an optimal temperature distribution in the
metal block.
Place the temperature dry-well calibrator on a clean
surface so that the fan on the bottom is not blocked and
sufficient fresh air can be drawn in.
Insufficient ventilation can lead to damage to the
calibrator.
EN
11

Publicidad

Tabla de contenido
loading

Tabla de contenido