Thank you for purchasing a MN93, MN93A, C193 or PAC93 AC current clamp or
an AmpFLEX A193 or A196 flexible current sensor. This current sensor should
only be used only with C.A 82XX, C.A 83XX and PEL 1XX analysers.
For best service from your current sensor:
„ read these operating instructions carefully,
„ comply with the precautions for use.
WARNING, risk of DANGER! The operator must refer to these instruc‑
tions whenever this danger symbol appears.
Current sensor protected by double insulation.
Application or withdrawal authorized on conductors carrying dangerous
voltages. Type A current sensor as per IEC 61010‑2‑032.
Must not be applied to or removed from conductors at dangerous
voltages. Type B current sensor as per IEC 61010‑2‑032.
The arrow indicates the phase of the current.
Battery.
The CE marking guarantees conformity with European directives, in
particular LVD and EMC.
The rubbish bin with a line through it indicates that, in the European
Union, the product must undergo selective disposal in compliance with
Directive WEEE 2002/96/EC. This equipment must not be treated as
household waste.
Definition of measurement categories:
„ Measurement category IV corresponds to measurements taken at the source of
low‑voltage installations.
Example: power feeders, counters and protection devices.
„ Measurement category III corresponds to measurements on building installations.
Example: distribution panel, circuit‑breakers, machines or fixed industrial devices.
„ Measurement category II corresponds to measurements taken on circuits directly
connected to low‑voltage installations.
Example: power supply to electro‑domestic devices and portable tools.