Aerosol Data According To Iso 27427 - Pari LC PLUS Instrucciones De Use

Ocultar thumbs Ver también para LC PLUS:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 69
6.2 Aerosol data according to ISO 27427
The aerosol characteristics presented in these instructions for
use were determined in accordance with ISO 27427 using Sal-
butamol. If other solutions or suspensions are used for
nebulisation, the aerosol characteristics may differ from the val-
ues shown (particularly if they have greater viscosity).
The following data is based on tests according to a standard
which takes adult breathing patterns as a basis. Therefore,
these figures will probably differ from corresponding figures
that were calculated for populations of children and infants.
Nozzle insert (red)
7
MMAD [µm]
8
GSD
Respirable fraction
[% < 5 µm]
Aerosol fraction
[% < 2 µm]
Aerosol fraction
[% > 2 µm < 5 µm]
Aerosol fraction
[% > 5 µm]
Aerosol output [ml] 0.44
Aerosol output rate
[ml/min]
Residual volume
[ml] (gravimetric)
6) Operation with PARI COMPACT2 compressor (Type 152).
7) MMAD = Mass Median Aerodynamic Diameter
8) GSD = Geometric Standard Deviation
– 24 –
Minimum com-
pressor flow
(3 l/min –
0.6 bar)
3.3
2.03
71.8
23.6
48.2
28.2
0.05
0.99
Nominal com-
pressor flow
(4 l/min –
6
1.2 bar)
2.7
2.17
79.2
35.0
44.2
20.8
0.49
0.07
0.94
Maximum com-
pressor flow
(6 l/min –
1.9 bar)
2.0
2.23
86.3
49.4
36.9
13.7
0.34
0.08
0.99