Storage; Troubleshooting; Technical Data; General Nebuliser Data - Pari LC PLUS Instrucciones De Use

Ocultar thumbs Ver también para LC PLUS:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 69

4.7 Storage

Store this product as described below:
• Wrap all individual components in a clean, lint-free cloth
(e.g., a tea towel).
• Store this product in a dry, dust-free place.
5

TROUBLESHOOTING

Fault
No aerosol is
coming out of
the nebuliser.
6

TECHNICAL DATA

6.1 General nebuliser data

5
Size
5
Weight
Operating gases
Minimum compressor flow
Minimum operating pressure
Maximum compressor flow
Maximum operating pressure
Minimum fill volume
Maximum fill volume
5) Without mouthpiece; unfilled.
– 23 –
Possible cause
The nebuliser
nozzle insert is
blocked.
The connection
tubing is not con-
nected properly.
The connection
tubing is leaking.
Remedy
Clean the nebuliser.
Check that the tubing connect-
ors are connected firmly to the
compressor and the nebuliser.
Replace the connection tubing.
10 cm × 9.5 cm × 4.5 cm
27.5 g to 29.5 g
Air
3.0 l/min.
0.5 bar / 50 kPa
6.0 l/min.
2.0 bar / 200 kPa
2 ml
8 ml; PARI LC STAR: 6 ml

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Lc plus babyLc star

Tabla de contenido